When entering a confined space, you must pay attention to the choice of combustible gas sensor
Publication Date:2020/12/18Gas detectors are used more in industrial production, and corresponding safety protection articles are used according to their detection results. What are the common failures of gas detectors? How to solve the common failures of gas detectors?
Failure 1. Low-concentration gas cannot be detected.
Solution ::
1. Check whether the air pump of the gas detector is working properly, and block the air inlet with your fingers for 5 seconds. If it feels normal, it has obvious attraction. No attraction, check whether the air inlet is blocked.

2, zero point calibration with nitrogen or calibration in clean air, check after calibration
3.After calibrating the zero point, if the measured gas cannot be detected, the factory settings of the gas detector need to be restored
4. the above steps have been operated, but they cannot be detected. it is necessary to confirm whether there is the measured gas on site, or the concentration of the measured gas is really very low, which cannot be detected if it is lower than the minimum detection accuracy of the gas sensor.
Fault 2. There is no measured gas in the air, but the value changes greatly or jumps.
solution:
1. the short-time zero point variation range is less than 1% of the maximum range, and the long-time drift of unmeasured gas is less than 2% of the maximum range. when exceeding this range, it is necessary to confirm whether there is measured gas on site, and the temperature and humidity in the air vary greatly, the value is unstable
2. confirm whether the gas detector carries out zero point calibration or target point calibration operation. if there is detected gas, the zero point calibration operation may not be able to detect low concentration gas. if there is detected gas, the target point calibration is carried out, but the calibrated concentration value is inconsistent with the actual concentration value, the value of the gas detector changes greatly or the detected value becomes smaller
4. If the problem cannot be solved, confirm whether the gas detector impacts the gas sensor through high-concentration gas or high-concentration gas.If the gas sensor is impacted, after the gas sensor is started and aged for 24 hours, if the value is unstable, the gas sensor may be impacted and damaged, and the gas sensor needs to be replaced.
Failure 3. The check is incorrect.
solution ::
1. confirm whether the gas concentration on site is correct, and there is a big gap between the theoretical value and the actual value. calibrate the gas detector by standard gas to ensure the correctness of detection, or send it to a third-party metering organization for calibration
2. if the gas sensor is used for a long time, the measured value may have errors. It is necessary to confirm with the manufacturer whether the gas sensor can continue to be used. If the sensor itself is close to the service life, it can be used normally in a short time even after re-labeling, but the measured value of the gas detector will drift and the detection will be incorrect. It is recommended to replace the gas sensor.
Fault 4, the value is 0 or the alarm value is not reached in the air.
solution:
1. check whether the various alarm parameters of the gas detector have been modified
2.Check whether the alarm mode and alarm mode of the gas detector have been modified
3. check whether the alarm state of the gas detector is a concentration alarm or other fault alarm. the words A1 or A2 appear in the concentration alarm and the red light flashes
4. if the gas detector alarm is caused by human operation modification, it can be solved by restoring factory installation, fault alarm needs to be further checked for short circuit, open circuit, poor contact, sensor failure, etc., or return to the factory for inspection.
Related information
-
Yiyuntian Eranntex Portable 4-in-1 Gas Detectors: Enhancing On-Site Safety Management for Petrochemical Production Enterprises!
The petrochemical industry involves numerous production processes and complex operational scenarios. Areas such as plant facilitie...
2026-04-30 -
Yiyuntian Eranntex Invites You to the 2026 Shanghai Emergency Response and Disaster Reduction Expo
As spring draws to a close, Shanghai is set to host a major event. From May 7 to 9, 2026, the 5th Yangtze River Delta Internationa...
2026-04-29 -
Application of Yiyuntian gas detector: accurate control of gas concentration to avoid accidents
The explosion accident of Weifang Gaomi Chemical Plant once again reminds us with the lesson of blood that there is no room for sl...
2025-06-06 -
Dust explosion hidden danger? Yiyuntian industrial explosion-proof dust detector to save the field!
With its excellent technical performance, Yiyuntian industrial explosion-proof dust detector has become a key weapon to crack the ...
2025-06-04 -
Breaking the game: Yiyuntian online dust detector to resolve the dilemma of CGNPC dust monitoring
Recently, CGN has encountered unprecedented challenges when promoting the new online dust detector project. The particularity of t...
2025-06-03
















info@eranntexgas.com


13480931872